The Journal of International
Advanced Otology
Original Article

Acoustic Analysis of Used Tuning Forks

1.

Department of Audiology, Ankara University School of Medicine, Ankara, Turkey

2.

Department of Otorhinolaryngology, Gazi University School of Medicine, Ankara, Turkey

J Int Adv Otol 2017; 13: 239-242
DOI: 10.5152/iao.2017.2745
Read: 2025 Downloads: 978 Published: 03 September 2019

Abstract

OBJECTIVE: In this study, we evaluated used aluminum tuning forks (TFs) for fundamental frequencies (FF), overtones, and decay times.

 

MATERIALS and METHODS: In total, 15 used (1 C1, 11 C2, and 3 C3) and 1 unused (C2) TFs were tuned, and the recorded sound data were analyzed using the Praat sound analysis program.

 

RESULTS: It was found that FFs of the recorded sounds produced by the used C2-TFs presented a high variability from 0.19% to 74.15% from the assumed FFs, whereas this rate was smaller (1.49%) in the used C3-TFs. Further, decay times of the used C2-TFs varied from 5.41 to 40.97 s.

 

 

CONCLUSION: This study, as the first of its kind in the literature, reported that some of the used aluminum TFs lost their physical properties that are important for clinical TF tests. It could be said that this is a phenomenon related to metal fatigue, which is common in aluminum products due to the cyclic load.

 

 

Cite this article as: Yüksel M, Kemaloğlu YK. Acoustic Analysis of Used Tuning Forks. J Int Adv Otol 2017; 13: 239-42. 